Repair cases

(57)
China OLYMPUS CF-2T160L scope for repair factory

OLYMPUS CF-2T160L scope for repair

OLYMPUS CF-2T160L scope, SN:210**** Fault: 1. With image. 2. 10% light bundle broken. 3. Insert tube is ok. 4. Biopsy tube abrasion. 5. Angle tied, bending section transformation. 6. Wires, helix tube abrasion, ... Read More
2024-03-19 09:34:45
China OLYMPUS GIF-1TQ160 gastroscope for repair factory

OLYMPUS GIF-1TQ160 gastroscope for repair

OLYMPUS GIF-1TQ160 gastroscope , SN:270**** Fault: 1.With image, but CCD with broken. 2.Light bundle broken completely. 3.Insert tube abrasion. 4. No biopsy tube. 5. No air&water tube, no water supply tube. 6. ... Read More
2024-03-19 09:34:45
China OLYMPUS GIF-Q180 gastroscope for repair factory

OLYMPUS GIF-Q180 gastroscope for repair

OLYMPUS GIF-Q180 gastroscope, SN: 270**** Fault: 1.image definition not enough, CCD aging. 2.30% light bundle broken. 3. Insert tube aging and wrinkle. 4.Biopsy tube abrasion. 5. Bending section bited, ... Read More
2024-03-19 09:34:45
China Siemens Acuson 18L6 probe for repair factory

Siemens Acuson 18L6 probe for repair

Siemens Acuson 18L6 probe for repair, SN: 335***** Fault: 1.Probe lens broken and soften; 2.Probe cable broken; 3.Scanhead aged Solution: Repair cable and replace probe lens Read More
2024-03-19 09:34:45
China Siemens Acuson 6C1 probe for repair factory

Siemens Acuson 6C1 probe for repair

Siemens Acuson 6C1 probe for repair, SN: 643***** Fault: 1. Housing damaged; 2. Probe cable broken; 3. Image with shadow and disturbance; 4. Scanhead aged Solution: Replace scanhead and housing Read More
2024-03-19 09:34:45
China OLYMPUS BF-XP160F fiber scope for repair factory

OLYMPUS BF-XP160F fiber scope for repair

OLYMPUS BF-XP160F fiber scope, SN:280**** Fault: 1. Image degumming, serious color, CCD aging, a few black and gray spots; 2. 20% of the light bundle is broken; 3. Insert tube aging slightly; 4. Biopsy tube ... Read More
2024-03-19 09:34:45
China STORZ 13820pks Gastroscope for Repair factory

STORZ 13820pks Gastroscope for Repair

STORZ 13820pks Gastroscope for Repair Fault: 1.Rubber leakage; 2.Wire abrasion; 3. no up angle, other angle lack of; Solution: 1. Replace the wires, coil pipe, rubber; 2.Adjust angle Read More
2024-03-19 09:34:45
China Insighters  IS3-F laryngoscope for repair, sn:CAA19**** factory

Insighters IS3-F laryngoscope for repair, sn:CAA19****

Insighters IS3-F laryngoscope for repair, sn:CAA19**** Fault: 1. Image blank screen; 2. Insert tube abrasion 3. Bending section transformation; 4. Distal head transformation; Suggestion: Replace CCD, insert ... Read More
2024-03-19 09:34:45
China GE 4D3C-L 4D Convex Transducer,(SN:2616****) factory

GE 4D3C-L 4D Convex Transducer,(SN:2616****)

GE 4D3C-L 4D Convex Transducer,(SN:2616****) Fault: 1.Probe dome abrasion & leakage, 2. Dome tube damaged, 3. 4D function not work, 4. Electric motor damaged. Suggestion: Replace the dome, dome tube; repair ... Read More
2024-03-19 09:34:45
China STORZ 13925PKS Colonoscope, for repair,SN: 06***. factory

STORZ 13925PKS Colonoscope, for repair,SN: 06***.

STORZ 13925PKS Colonoscope, for repair,SN: 06***. Fault: Grip leakage, no angle. Solution: Replace wires, coil pipe,mesh, seal ring for grip and rubber. Adjust the angle. Replace angle part from the second ... Read More
2024-03-19 09:34:45