| Brand: | SonoScape | Model: | EC-500T |
|---|---|---|---|
| Inspection Results: | 1. Image Display Present; 2. Approximately 20% Fibers Broken; 3. Insertion Tube In Acceptable Condition; 4. Insufficient Angle, Wear On Snake Bone, Wire, And Helix Tube; 5. Light Guide Tube Worn; 6. Poor Sealing Of Connector Housing, Leakage; 7. Poor | Maintenance Plan: | Replace The Following Components: Insertion Tube, Biopsy Tube, Water/air Tube, Snake Bone, Wire, Helix Tube, Light Guide Tube, And Rubber Seal. CMOS CCD Not Covered Under Warranty. |
SonoScape EC-500T Electronic Colonoscope For Repair
► Brief Description
Inspection Results:
1. Image display present;
2. Approximately 20% fibers broken;
3. Insertion tube in acceptable condition;
4. Insufficient angle, wear on snake bone, wire, and helix tube;
5. Light guide tube worn;
6. Poor sealing of connector housing, leakage;
7. Poor sealing of S cover, leakage.
Maintenance plan:
Replace the following components: insertion tube, biopsy tube, water/air tube, snake bone, wire, helix tube, light guide tube, and rubber seal. CMOS CCD not covered under warranty.
► Common issues with SonoScape EC-500T
System Communication Faults
►FAQ
A: Foggy or blurred images are usually caused by moisture ingress behind the distal lens, lens delamination, or internal condensation.
Perform a routine leakage test to check for structural breaches.
If moisture is found inside, immediately stop use to prevent corrosion of the CCD sensor.
Verification: The repair technician must open the distal tip, clean and dry the optical components in a sterile environment, and reseal the unit using medical-grade epoxy.
A: Black spots are typically caused by broken glass fibers in the light guide bundle, dust on the CCD surface, or damaged pixel elements on the image sensor.
Verification: Inspect the light guide bundle using a high-intensity light source. If dead pixels or fixed black spots persist on the monitor regardless of external lighting, the CCD module or objective lens assembly must be replaced by a certified technician.
If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you solve the issue.
► About Us
JDP is the global provider of medical equipment and medical equipment repair service, as well as offering spare parts. We focus on repair and repair training for monitor, Flexible & Rigid Endoscope, ultrasound, camera system, dynamical system and surgical instrument.
Shipping and packing
All products will be tested and sterilized for free, perfectly packed to ensure the safety of the goods, strictly in accordance with international export rules. We send to worldwide via EMS, DHL, FEDEX or other express, based on your choice.
Contact Person: Mr. Maud
Tel: 13922729804