Product Details:

Model Number: BF-XP160F

Payment & Shipping Terms:

Minimum Order Quantity: 1
Payment Terms: T/T, Western Union
Detailed Product Description
Condition: Repaired

OLYMPUS BF-XP160F fiber scope, SN:280****

 

Fault:

1. Image degumming, serious color, CCD aging, a few black and gray spots;

2. 20% of the light bundle is broken;

3. Insert tube aging slightly;

4. Biopsy tube wear, not smooth;

5. Bending section, steel wire, screw tube wear;

 

Maintenance plan:

Suggestion: replace CCD, image beam, beam, insert tube, pliers tube, snake bone, steel wire, spiral tube, rubber.

 

Repair cost: $****

 